Detonation gun sprayed Al2O3–13TiO2coatings for biomedical applications
, , Chava K, Joshi S.V.
Published in Informa UK Limited
2014
Volume: 30
   
Issue: 4
Pages: 229 - 236
Abstract
The present work deals with the investigation of the corrosion and sliding wear behaviour of detonation gun sprayed (DG) conventional and nanostructured Al2O3-13TiO2 coatings deposited on a biomedical grade Ti-13Nb-13Zr alloy. The microstructure and phase composition of the coatings were characterised by scanning electron microscopy (SEM) and X-ray diffraction (XRD). Hardness measurements were carried out using a Vickers hardness testing machine. The experimental results suggested that the nano structured Al2O3-13TiO2 coating exhibited 43 and 33% increase respectively in corrosion and wear resistances compared to the conventional Al2O3-13TiO2 coating. This improvement in the above properties of nanostructured Al2O 3-13TiO2 coating is due to the presence of larger volume fraction of nanosized particles and lower porosity attained by spraying using DG. © 2014 Institute of Materials, Minerals and Mining.
About the journal
JournalSurface Engineering
PublisherInforma UK Limited
ISSN0267-0844
Open Access0